Products
FROM CRITICAL EQUIPMENT TO
TOTAL PROCESS SOLUTION
Filling Line System
Depyrogenation Tunnel

The depyrogenation tunnel designed to reduce the endotoxin at least 3 log and reduce the bacteria at least 6 log.

Highlights

• Modular design, each is one independent unit.

• The inner chamber is fully welded, and the expansion joint is used between the chambers to prevent cracking.

• Heat penetration effect FH≥30min,based on T0=250℃,Z=46.4℃.

• Endotoxin ≤0.1EU, ensure endotoxin at least 3log drop and bacteria at least 6log log drop.

• Temperature uniformity can be achieved ±5℃, air velocity uniformity can be achieved ±20%

• △P≥10Pa between tunnel inside and cleaning room is possible.

• Negative pressure seal for HEPA

• The fan driver adopts self-cooling mode without additional cooling medium

• Equipped with air pressure balance system  in order to ensure the pressure is stable.

• CIP is option for conveyor belt.

• SIP is option for cooling zone.

• The sealing door is option to separate completely the cooling zone and aseptic area.

• Wide vial application from 2ml ~ 100ml.

Main Technical Data

Model APV2SW050 APV2SW100 APV2SW200 APV2SW300 APV2SW400 APV2SW500
Container Range (ml) 2-100 2-100 2-100 2-100 2-100 2-100
Capacity
(VPM)
10ml
(φ22mm)
50 100 200 300 400 500
Purified Water Supply (m³/h) 0.3 0.4 0.6 0.6 0.6 0.8
Chilled Water Supply (m³/h) 1.5-2.5 4-6 5-7 7-10 7-10 10-14
Compressed Air for control Supply(m³/h) 3 3 3 3 3 3
Process Compressed Air Supply(m³/h) 15 15 15 15 15 18
Air Supply of Infeed Zone (m³/h) 900-1300 1400-1800 1500-2000 1500-2000 1500-2000 1760-2350
Air Supply of Heating Zone (m³/h) 100-200 200-400 300-500 300-500 300-500 470-705
Normal Power (kw) 40 72 85 108 130 148
Dimension L*W*H (mm) 2545*1655*2407 3168*1825*2407 3808*2075*2464 4117*2075*2464 5273*2075*2464 5467*2225*2464
Weight (kg) 3500 4000 4800 5500 7000 8500